TY - JOUR
T1 - A 12-channel flexible receiver coil for accelerated tongue imaging
AU - Voskuilen, L.
AU - de Heer, P.
AU - van der Molen, L.
AU - Balm, A.J.M.
AU - van der Heijden, F.
AU - Strijkers, G.J.
AU - Smeele, L.E.
AU - Nederveen, A.J.
PY - 2020/8/1
Y1 - 2020/8/1
N2 - © 2020, The Author(s).Objective: MRI of the tongue requires acceleration to minimise motion artefacts and to facilitate real-time imaging of swallowing. To accelerate tongue MRI, we designed a dedicated flexible receiver coil. Materials and methods: We designed a flexible 12-channel receiver coil for tongue MRI at 3T and compared it to a conventional head-and-neck coil regarding SNR and g-factor. Furthermore, two accelerated imaging techniques were evaluated using both coils: multiband (MB) diffusion-tensor imaging (DTI) and real-time MRI of swallowing. Results: The flexible coil had significantly higher SNR in the anterior (2.1 times higher, P = 0.002) and posterior (2.0 times higher, P < 0.001) parts of the tongue, while the g-factor was lower at higher acceleration. Unlike for the flexible coil, the apparent diffusion coefficient (P = 0.001) and fractional anisotropy (P = 0.008) deteriorated significantly while using the conventional coil after accelerating DTI with MB. The image quality of real-time MRI of swallowing was significantly better for hyoid elevation (P = 0.029) using the flexible coil. Conclusion: Facilitated by higher SNR and lower g-factor values, our flexible tongue coil allows faster imaging, which was successfully demonstrated in MB DTI and real-time MRI of swallowing.
AB - © 2020, The Author(s).Objective: MRI of the tongue requires acceleration to minimise motion artefacts and to facilitate real-time imaging of swallowing. To accelerate tongue MRI, we designed a dedicated flexible receiver coil. Materials and methods: We designed a flexible 12-channel receiver coil for tongue MRI at 3T and compared it to a conventional head-and-neck coil regarding SNR and g-factor. Furthermore, two accelerated imaging techniques were evaluated using both coils: multiband (MB) diffusion-tensor imaging (DTI) and real-time MRI of swallowing. Results: The flexible coil had significantly higher SNR in the anterior (2.1 times higher, P = 0.002) and posterior (2.0 times higher, P < 0.001) parts of the tongue, while the g-factor was lower at higher acceleration. Unlike for the flexible coil, the apparent diffusion coefficient (P = 0.001) and fractional anisotropy (P = 0.008) deteriorated significantly while using the conventional coil after accelerating DTI with MB. The image quality of real-time MRI of swallowing was significantly better for hyoid elevation (P = 0.029) using the flexible coil. Conclusion: Facilitated by higher SNR and lower g-factor values, our flexible tongue coil allows faster imaging, which was successfully demonstrated in MB DTI and real-time MRI of swallowing.
U2 - 10.1007/s10334-019-00824-5
DO - 10.1007/s10334-019-00824-5
M3 - Article
SN - 0968-5243
VL - 33
SP - 581
EP - 590
JO - Magnetic Resonance Materials in Physics, Biology and Medicine
JF - Magnetic Resonance Materials in Physics, Biology and Medicine
IS - 4
ER -