TY - JOUR
T1 - Cavity ring-down spectroscopy for detection in liquid chromatography at UV wavelengths using standard cuvettes in a normal incidence geometry
AU - van der Sneppen, L.
AU - Ariese, F.
AU - Gooijer, C.
AU - Ubachs, W.M.G.
PY - 2007
Y1 - 2007
N2 - Liquid chromatography (LC) with cavity ring-down spectroscopy (CRDS) detection, using flow cuvettes (put under normal incidence inside the ring-down cavity), is demonstrated. Fresnel reflections are maintained within the capture range of a stable cavity of 4 cm length. This method circumvents the need for specific Brewster's angles and possible mirror degradation is avoided. The flow cuvettes are commercially available at low cost. At 355 nm (the frequency-tripled output of a Nd:YAG laser), the system surpasses the performance of conventional absorbance detectors; the baseline noise was 1.3 × 10
AB - Liquid chromatography (LC) with cavity ring-down spectroscopy (CRDS) detection, using flow cuvettes (put under normal incidence inside the ring-down cavity), is demonstrated. Fresnel reflections are maintained within the capture range of a stable cavity of 4 cm length. This method circumvents the need for specific Brewster's angles and possible mirror degradation is avoided. The flow cuvettes are commercially available at low cost. At 355 nm (the frequency-tripled output of a Nd:YAG laser), the system surpasses the performance of conventional absorbance detectors; the baseline noise was 1.3 × 10
U2 - 10.1016/j.chroma.2007.03.028
DO - 10.1016/j.chroma.2007.03.028
M3 - Article
SN - 0021-9673
VL - 1148
SP - 184
EP - 188
JO - Journal of Chromatography A
JF - Journal of Chromatography A
IS - 2
ER -