Heat flux systems for body core temperature assessment during exercise

  • Hein A.M. Daanen*
  • , Veerle Kohlen
  • , Lennart P.J. Teunissen
  • *Corresponding author for this work

Research output: Contribution to JournalArticleAcademicpeer-review

367 Downloads (Pure)

Abstract

Heat flux systems are increasingly used to assess core body temperature. However, validation of multiple systems is scarce. Therefore, an experiment was performed in which three commercially available heat flux systems (3 M, Medisim and Core) were compared to rectal temperature (Tre). Five females and four males performed exercise in a climate chamber set at 18 °C/50% relative humidity until exhaustion. Exercise duration was 36.3 ± 5.6 min (mean ± standard deviation). Tre in rest was 37.2 ± 0.3 °C. Medisim's-values were lower than Tre (36.9 ± 0.4 °C, p < 0.05); 3 M (37.2 ± 0.1 °C) and Core's (37.4 ± 0.3 °C) did not differ from Tre. Maximal temperatures after exercise were 38.4 ± 0.2 °C (Tre), 38.0 ± 0.4 °C (3 M), 38.8 ± 0.3 °C (Medisim) and 38.6 ± 0.3 °C (Core); Medisim was significantly higher than Tre (p < 0.05). The temperature profiles of the heat flux systems during exercise differed to varying degree from the rectal profiles; the Medisim system showed a faster increase during exercise than Tre (0.48 ± 0.25 °C in 20 min, p < 0.05), the Core system tended to show a systematic overestimation during the entire exercise period and the 3 M system showed large errors at the end of exercise, likely due to sweat entering the sensor. Therefore, the interpretation of heat flux sensor values as core body temperature estimates should be done with care; more research is required to elucidate the physiological significance of the generated temperature values.

Original languageEnglish
Article number103480
Pages (from-to)1-6
Number of pages6
JournalJournal of Thermal Biology
Volume112
Early online date14 Jan 2023
DOIs
Publication statusPublished - Feb 2023

Bibliographical note

Funding Information:
Surgical Company is acknowledged for providing financial support for the study. This company had no involvement in study design, in the collection, analysis and interpretation of data and in the writing of the report. The company approved the decision to submit the article for publication.

Funding Information:
Surgical Company is acknowledged for providing financial support for the study. This company had no involvement in study design, in the collection, analysis and interpretation of data and in the writing of the report. The company approved the decision to submit the article for publication.

Publisher Copyright:
© 2023

Funding

Surgical Company is acknowledged for providing financial support for the study. This company had no involvement in study design, in the collection, analysis and interpretation of data and in the writing of the report. The company approved the decision to submit the article for publication. Surgical Company is acknowledged for providing financial support for the study. This company had no involvement in study design, in the collection, analysis and interpretation of data and in the writing of the report. The company approved the decision to submit the article for publication.

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 7 - Affordable and Clean Energy
    SDG 7 Affordable and Clean Energy

Keywords

  • Accuracy
  • Rectal temperature
  • Skin temperature
  • Thermometry

Fingerprint

Dive into the research topics of 'Heat flux systems for body core temperature assessment during exercise'. Together they form a unique fingerprint.

Cite this