TY - JOUR
T1 - MP2RAGEME: T1, T2*, and QSM mapping in one sequence at 7 tesla
AU - Caan, M.W.
AU - Bazin, P.L.
AU - Marques, J.P.
AU - de Hollander, G.
AU - Dumoulin, S.O.
AU - van der Zwaag, W.
PY - 2018
Y1 - 2018
N2 - Quantitative magnetic resonance imaging generates images of meaningful physical or chemical variables measured in physical units that allow quantitative comparisons between tissue regions and among subjects scanned at the same or different sites. Here, we show that we can acquire quantitative T
1 , T
2
* , and quantitative susceptibility mapping (QSM) information in a single acquisition, using a multi-echo (ME) extension of the second gradient-echo image of the MP2RAGE sequence. This combination is called MP2RAGE ME, or MP2RAGEME. The simultaneous acquisition results in large time savings, perfectly coregistered data, and minimal image quality differences compared to separately acquired data. Following a correction for residual transmit B
1
+ -sensitivity, quantitative T
1 , T
2
* , and QSM values were in excellent agreement with those obtained from separately acquired, also B
1
+ -corrected, MP2RAGE data and ME gradient echo data. The quantitative values from reference regions of interests were also in very good correspondence with literature values. From the MP2RAGEME data, we further derived a multiparametric cortical parcellation, as well as a combined arterial and venous map. In sum, our MP2RAGEME sequence has the benefit in large time savings, perfectly coregistered data and minor image quality differences.
AB - Quantitative magnetic resonance imaging generates images of meaningful physical or chemical variables measured in physical units that allow quantitative comparisons between tissue regions and among subjects scanned at the same or different sites. Here, we show that we can acquire quantitative T
1 , T
2
* , and quantitative susceptibility mapping (QSM) information in a single acquisition, using a multi-echo (ME) extension of the second gradient-echo image of the MP2RAGE sequence. This combination is called MP2RAGE ME, or MP2RAGEME. The simultaneous acquisition results in large time savings, perfectly coregistered data, and minimal image quality differences compared to separately acquired data. Following a correction for residual transmit B
1
+ -sensitivity, quantitative T
1 , T
2
* , and QSM values were in excellent agreement with those obtained from separately acquired, also B
1
+ -corrected, MP2RAGE data and ME gradient echo data. The quantitative values from reference regions of interests were also in very good correspondence with literature values. From the MP2RAGEME data, we further derived a multiparametric cortical parcellation, as well as a combined arterial and venous map. In sum, our MP2RAGEME sequence has the benefit in large time savings, perfectly coregistered data and minor image quality differences.
UR - http://www.scopus.com/inward/record.url?scp=85058465269&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85058465269&partnerID=8YFLogxK
U2 - 10.1002/hbm.24490
DO - 10.1002/hbm.24490
M3 - Article
SN - 1065-9471
VL - 40
SP - 1786
EP - 1798
JO - Human Brain Mapping