Abstract
An on-line liquid chromatography-immunochemical detection (LC-ICD) system for the quantification of cytokines in cell extracts has been developed using a post-column continuous-flow reaction detection system using fluorescence labelled antibodies. Cytokines eluting from the micro-HPLC column react with antibodies to form fluorescent complexes. In a second step the excess of free antibody is trapped on a cytokine bound support prior to fluorescence detection. The concentration detection limit of the flow injection-ICD system was 50 pM (20 μl injection volume) for interleukine 4 (IL-4). An absolute detection limit of 1 fmol was obtained for IL-4. Similar to ICD systems for small non-protein analytes developed earlier, reaction times were in the order of 1 minute. The immobilised cytokine affinity columns can easily be regenerated and used for months. The present ICD system for interleukine 4, 6, 8 and 10 was coupled to ion exchange-, size exclusion- and reversed phase chromatography. Important parameters (reaction times, reaction conditions) were investigated to get a better understanding of post-column ICD systems for macromolecules. © 2001 Elsevier Science B.V. All rights reserved.
Original language | English |
---|---|
Pages (from-to) | 975-85 |
Journal | Journal of Pharmaceutical and Biomedical Analysis |
Volume | 26 |
Issue number | 5-6 |
DOIs | |
Publication status | Published - 2001 |