Validation of Photoplethysmography Using a Mobile Phone Application for the Assessment of Heart Rate Variability in the Context of Heart Rate Variability-Biofeedback

Research output: Contribution to JournalArticleAcademicpeer-review

88 Downloads (Pure)

Abstract

OBJECTIVE: Heart rate variability-biofeedback (HRV-BF) is an effective intervention to reduce stress and anxiety and requires accurate measures of real-time HRV. HRV can be measured through photoplethysmography (PPG) using the camera of a mobile phone. No studies have directly compared HRV-BF supported through PPG against classical electrocardiogram (ECG). The current study aimed to validate PPG HRV measurements during HRV-BF against ECG. METHODS: Fifty-seven healthy participants (70% women) with a mean (standard deviation) age of 26.70 (9.86) years received HRV-BF in the laboratory. Participants filled out questionnaires and performed five times a 5-minute diaphragmatic breathing exercise at different paces (range, ~6.5 to ~4.5 breaths/min). Four HRV indices obtained through PPG, using the Happitech software development kit, and ECG, using the validated NeXus apparatus, were calculated and compared: RMSSD, pNN50, LFpower, and HFpower. Resonance frequency (i.e., optimal breathing pace) was also compared between methods. RESULTS: All intraclass correlation coefficient values of the five different breathing paces were "near perfect" (>0.90) for all HRV indices: lnRMSSD, lnpNN50, lnLFpower, and lnHFpower. All Bland-Altman analyses (with just three incidental exceptions) showed good interchangeability of PPG- and ECG-derived HRV indices. No systematic evidence for proportional bias was found for any of the HRV indices. In addition, correspondence in resonance frequency detection was good with 76.6% agreement between PPG and ECG. CONCLUSIONS: PPG is a potentially reliable and valid method for the assessment of HRV. PPG is a promising replacement of ECG assessment to measure resonance frequency during HRV-BF.

Original languageEnglish
Pages (from-to)568-576
Number of pages9
JournalPsychosomatic medicine
Volume85
Issue number7
Early online date1 Sept 2023
DOIs
Publication statusPublished - Sept 2023

Bibliographical note

Publisher Copyright:
Copyright © 2023 by the American Psychosomatic Society.

Funding

Source of Funding and Conflicts of Interest: This work was funded by ZonMW (project number 543003104) awarded to A.C.H. The authors report no conflicts of interest.

FundersFunder number
ZonMw543003104

    Fingerprint

    Dive into the research topics of 'Validation of Photoplethysmography Using a Mobile Phone Application for the Assessment of Heart Rate Variability in the Context of Heart Rate Variability-Biofeedback'. Together they form a unique fingerprint.

    Cite this